Fully automatic operation from weighing individual capsule to printing report with statistical data.
Precision weighing balance with 1mg accuracy and repeatability.
Interfacing system to switch off the machine, if capsules are beyond the set limits.
Completely enclosed & sturdy design, which will not transfer the vibration of the machine to the weighing balance.
Provision for recording graph, batch data, date & time, statistics result & operator detail etc.
Either more than 1 or no capsules recorded, Programmer will not consider the weight & will not print.
Provision of viewing the reading sample being weighed on the monitor.
Provision of viewing the result on Tower indicator.
Overfilling is a waste and underfilling is unacceptable. The solution for all your laboratory needs is CITIZENs line of SQC systems, which control and regulate all aspects of the filling process.
Technical Specifications
Parameter/Model
Standerd Weight Subarea
Hopper capacity
30 capsules of 00
Capsule Sizes
00 to 05
Weighing balance details Model Supplier
Citizen CY 220, CY 360 Citizen Scales INC
Weighing capacity
220 gm Max. / 360 gm Max.
Least count
1 mg
Accuracy
+ 1 mg
Repeatability
+ 1 mg
Zero setting
Automatic by tare in the balance, before every sample
Limit setting
Programmable, using key board, + % of upper & lower limits
Interface
RS 232 C, BI directional, through controller
Ambient
5 to 30 deg. C, And R.H. 25 to 65%
Mains power
A/C Adapter 230V or 115V +/- 20% 50-60 Hz
Data print out
Lexmark Deskject Printer
Capsule feeding
24 VAC Electromagnetic vibrator with amplitude adjustable Pot Capsule feeding
Settable parameters
Sample Size (n): (maximum 30)
Set Weight: Batch No.: Tolerance Individual: Tolerance Average: Time / count base sampling: Sampling time: Sampling stroke: